821027: AMPx-SDA User Guide
© 2012 Wohler Technologies, Inc. All rights reserved.
15
AMPx-SDA User Guide
Rear Panel Connectors
Line Level Calibration: The unit is calibrated at the factory.
To recalibrate:
1) Turn on the power.
2) Apply a reference level (nominal 0) signal to all four
channels.
3) Make sure the Gain Calibration DIP sections (S2, S3) are set to
the nearest level (i.e., 0, +4, +6 or +8).
4) Place section S1 of the DIP switch in the DOWN position.
5) Wait 10 seconds. The unit will remove the previous calibration.
The unit will make sure that all four channels are within ±4 dB
of nominal zero. If all four channels are within this range
then the new calibration will be applied. If any channel is
outside this range then no new calibration will be
applied.
6) Place section S1 of the DIP switch in the UP position and
return unit to service. If you want to calibrate again, turn
off the power to the unit and repeat steps 1 through 6.
See Figure 1–8 on page 14 for settings.
Meter Gain Calibration: DIP switch sections S2 and S3
determine the Gain Calibration, which adjusts the level of the
input signal and the resultant "0" level displayed on the LED
bargraphs. Factory setting is +4 dB. See Figure 1–8 on page 14
for all settings.
Bar Graph Display Mode: DIP switch sections S4 and S5
determine how peak levels are displayed for the four
associated meters on the front panel. S4 selects either the PPM
level or a new auto-reset Peak Hold of the VU level (not the
PPM value). When S3 is set for VU Peak, S5 sets the Hold time
to either 3 or 10 seconds. If S4 is set for PPM Peak, S5
determines whether it is displayed in the original floating
PPM dot mode (no hold, continuous decay of 20 dB in 1.5
seconds) or as a 3 second auto-reset hold. The factory default
setting is PPM Peak - No Hold (continuous decay of 20 dB in
1.5 seconds). See Figure 1–8 on page 14 for all settings. Note
that section S6 on the DIP switch is not used.